Faced with a variety of pure water equipment, how should we choose and maintain pure water equipment?

Dec 5, 20235 min readVOLARDDA Engineering Team5 views

Faced with a variety of pure water equipment, how should we choose and maintain pure water equipment?

High pressure pump does not stop

  • The high pressure pump pressure is insufficient, can not reach the high pressure set pressure
  • The check valve is blocked and does not produce pure water
  • High pressure failure, unable to jump

When the water is full, the machine jumps repeatedly

  • Raw water pressure is insufficient
  • Relieve pressure from check valve
  • The high pressure switch or liquid level switch is faulty
  • The system has depressurized

The high-pressure pump is working, but it can't make water

  • The high pressure pump loses pressure
  • The inlet solenoid valve is faulty and cannot enter water (neither pure water nor wastewater)(Whether it is reversed).
  • The pre-filter element is blocked (no pure water wastewater or very small wastewater)
  • Check valve failure (waste water without pure water)
  • The automatic flushing solenoid valve fails and cannot be effectively closed (it has been in the flushing state)
  • The backflush solenoid valve cannot be closed if the computer case is faulty (it has been in the flushing state)
  • The RO membrane is blocked

Pure water flow is insufficient

  • The front filter element is blocked
  • The high pressure pump pressure is insufficient
  • The RO membrane is blocked
  • The waste water valve or waste water proportioner is too open
  • Retroactive carbon plugging
  • The pressure bucket has insufficient pressure or internal damage

High pressure pump does not start, can not make water

  • Check whether the power is off and the plug is plugged in
  • Check whether the low-voltage switch is faulty and cannot be turned on
  • Check whether the water pump and transformer are short-circuited, or the circuit connection is incorrect
  • Check whether the high pressure switch or water level controller is faulty and cannot be reset
  • Check whether the computer case is faulty

The high pressure pump stopped, but the waste water didn't stop

  • Solenoid valve failure, can not effectively cut off water
  • The computer box is faulty, and the solenoid valve cannot be closed (refers to the microcomputer type).
  • Pressure relief of check valve (small wastewater flow)